Top Seller
Fluke 742A-25 25ohm Resistance Standard
Elgin, IL


Condition info:


Fluke 742A-25 Features:
Small and rugged resistance calibrators
No oil or air baths required
18° to 28° C operating range
Six-month stability to 2.5 ppm
Supplied with temperature characterization
The Fluke 742A-25 Resistance Standard is a high accuracy working standard for precision, on-site resistance calibration. Its excellent temperature stability allows them to be used from 18°C to 28°C with typically less than 2 ppm degradation. Using the calibration table supplied with the standards, which lists corrections in 0.5°C increments, this uncertainty can be reduced to near zero. No cumbersome oil or air baths are required.
Because the Fluke 742A-25 Resistance Standard is small and rugged, it is easy to transport. Care has been taken to reduce resistance changes brought about by thermal and mechanical shock. Retrace (permanent shift in resistance) is typically less than 2 ppm after cycling between 0°C and 40°C.