Top Seller
Philips PM2811 30V/10A/60W, Power Supply
Location:
Elgin, IL
Model:
PM2811

Specifications

Condition:
used
Condition info:
Refurbished
Readback accuracy:
(referred to the output terminals)
Subcategory:
Dc power supply

Description

NIST Calibration Pricing - $115.00
Philips PM2811 Features:
OUTPUT module 30V/10A/60W
Output Voltage range 0 to 30 V
Output Current range 0 to 10 A
Output Power 60 W
Overvoltage protection range 2 to 32 V
Accuracy (at calibration temperature)
Output Voltage +/- 0.04 % of setting + 10 mV
Output Current +/- 0.1 % of setting + 50 mA
Over Voltage Protection +/- 0.5 % of setting + 150 mV
Temperature coefficient
Output Voltage (0.016 % of setting + 2 mV)/K
Output Current (0.05 % of setting + 2 mA)/K
Long term drift (in an 8 hour period)
Output Voltage 0.04 % of setting + 5 mV
Output Current 0.05 % of setting + 10 mA
Programming Resolution (12bit)
Output Voltage 7.5 mV/LSB
Output Current 2.5 mA/LSB
Over Voltage Protection 7.5 mV/LSB
Source Effect (for any line voltage change within rating)
Voltage +/- (0.01 % of setting + 2 mV)
Current +/- (0.02 % of setting + 2 mA)
Load Effect (load change 10 % to 90 % or 90 % to 10 % full load)
Constant Voltage +/- (0.01 % of setting + 5 mV)
Constant Current +/- (0.02 % of setting + 5 mA)
Load cross regulation Negligible for both voltage and current
Output Noise (PARD)
Constant Voltage pp/rms 15 mV/3 mV
Constant Current rms 10 mA
Switching Frequency approx. 20 kHz approx.
READBACK ACCURACY:(referred to the output terminals)
Accuracy (at calibration temperature)
Voltage +/- (0.05 % + 25 mV)
Current +/- (0.1 % + 30 mA)
Measuring Resolution (12 bit)
Voltage resolution to interface 7.5 mV
Current resolution to interface 2.5 mA
Voltage resolution to display 10 mV
Current resolution to display 10 mA
Max. Current at Reverse voltage 10 A
REMOTE SENSE > 0.25 Volt drop per lead (at open sense connections, the output voltage will not be more than 3 V above the programmed value)
DYNAMIC OPERATION ÆI=1 A between 1 and 10A between 0.5 and 5A between 1 and 10A
Settling band +/- 50 mV
dI/dt